PM600 Pressure Measurement Modules

PM600 Pressure Measurement Modules PM600 Pressure Measurement Modules

For 6270A Pressure Controller/Calibrator

Key features:

  • Fluke Calibration Quartz Reference Pressure Transducer (Q-RPT) technology provides precise measurement with long term stability
  • 0.01% reading measurement uncertainty from 30% to 100% of the modules’ span allows for extremely wide workload coverage
  • Onboard barometer included with absolute mode pressure modules, allowing them to be used in both absolute and gauge mode

Performance specifications

The performance specifications describe the complete instrumental uncertainty of the Product. The specifications include all relevant error components (linearity, hysteresis, repeatability, resolution, reference standard measurement uncertainty, 1-year drift, and temperature effects). The specifications are provided at a level of confidence of 95 %, k=2, normally distributed. Precision uncertainty includes linearity, hysteresis, repeatability, resolution, and temperature effects.

PM600 Modules measurement specifications

Specifications are valid from 15 °C to 35 °C.

Model Absolute Mode Range
(SI Units)
Absolute Mode Range
(Imperial Units)
Gauge Mode Range 3
(SI Units)
Gauge Mode Range
(Imperial Units)
1-Year
Instrumental Uncertainty
(% of reading or % FS, whichever is greater)
Precision Uncertainty
(% of reading or % FS, whichever is greater)
BRM600-BA100K4 70 kPa to 110 kPa 10 psi to 16 psi - - 0.01 % of reading 0.008 or 0.0024
PM600-BG15K - - -15 kPa to 15 kPa -60 inH2O to 60 inH2O 0.01 or 0.003 0.008 or 0.0024
PM600-G100K - - 0 kPa to 100 kPa 0 psi to 15 psi 0.01 or 0.003 0.008 or 0.0024
PM600-G200K - - 0 kPa to 200 kPa 0 psi to 30 psi 0.01 or 0.003 0.008 or 0.0024
PM600-A100K 6 kPa to 100 kPa 0.9 psi to 15 psi -94 kPa to 0 kPa -13.8 psi to 0 psi 0.01 or 0.003 1,3 0.008 or 0.0024
PM600-A200K 10 kPa to 200 kPa 1.5 psi to 30 psi -90 kPa to 100 kPa -13.2 psi to 15 psi 0.01 or 0.003 1,3 0.008 or 0.0024
PM600-A350K 10 kPa to 350 kPa 1.5 psi to 50 psi -90 kPa to 250 kPa -13.2 psi to 35 psi 0.01 or 0.003 1 0.008 or 0.0024
PM600-A700K 18 kPa to 700 kPa 2.6 psi to 100 psi -82 kPa to 700 kPa -12.1 psi to 100 psi 0.01 or 0.003 1 0.008 or 0.0024
PM600-A1.4M 0.035 MPa to 1.4 MPa 5 psi to 200 psi -0.065 MPa to 1.4 MPa -10 psi to 200 psi 0.01 or 0.003 1 0.008 or 0.0024
PM600-A2M 0.07 MPa to 2 MPa 10 psi to 300 psi -0.03 MPa to 2 MPa -5 psi to 300 psi 0.01 or 0.003 1 0.008 or 0.0024
PM600-A3.5M 0.07 MPa to 3.5 MPa 10 psi to 500 psi -0.03 MPa to 3.5 MPa -5 psi to 500 psi 0.01 or 0.003 1 0.008 or 0.0024
PM600-A7M ATM 2 to 7 MPa ATM 2 to 1000 psi 0 MPa to 7 MPa 0 psi to 1000 psi 0.01 or 0.003 1 0.008 or 0.0024
PM600-A10M ATM 2 to 10 MPa ATM 2 to 1500 psi 0 MPa to 10 MPa 0 psi to 1500 psi 0.01 or 0.003 1 0.008 or 0.0024
PM600-A14M ATM 2 to 14 MPa ATM 2 to 2000 psi 0 MPa to 14 MPa 0 psi to 2000 psi 0.01 or 0.003 1 0.008 or 0.0024
PM600-A20M ATM 2 to 20 MPa ATM 2 to 3000 psi 0 MPa to 20 MPa 0 psi to 3000 psi 0.01 or 0.003 1 0.008 or 0.0024
Notes

  1. For PM600s absolute mode modules used in absolute mode, root sum square (RSS) with 0.007 % of FS (reduced to k=1 by square root of 3).
formula
  1. ATM is any atmospheric pressure from 70 kPa to 110 kPa (10 psi to 16 psi).
  2. For absolute ranges used in gauge mode there is an additional uncertainty of ±7 Pa for dynamic barometric compensation. When combined with other uncertainties this changes the threshold uncertainty for the PM600-A100K to 0.008 % Span and for the PM600-A200K to 0.004 % Span.
  3. These modules can be used as a barometric reference module.