Hastings HFM-D-306 / HFC-D-308 High Capacity Flowmeters and Controllers

Hastings HFM-D-306 / HFC-D-308 High Capacity Flowmeters and Controllers

306/8-D Series mass flow instruments have flow ranges from 0-2500 slpm to 0-10,000 slm; use all metal seals on certain models, provide analog and digital outputs, and are constructed of stainless steel.

The Hastings 30x-D offer many enhanced features such as self diagnostics of the sensor and valve, auto zero feature, multi range and gas calibration storage, simple communications with control/logging software and totaliser.

± (0.5% of reading + 0.2% of full scale) Accuracy

Features
• +/- (0.5% of reading + 0.2% of Full Scale)
• Proven Reliability
• Range 2500 slpm to 10,000 slm(N2 Equivalent)
• Built in Totalizer
• Automatic Zeroing


Applications and Industries

  • Alternate Energy
  • Fuel Cell R&D
  • Custody Transfer
HFM-D-306A Meter HFC-D-308A Controller
Range 0 - 2500 slm to 0 - 10000 slm 0 - 2500 slm to 0 - 10000 slm
Accuracy + (0.5% of reading + 0.2% of full scale) + (0.5% of reading + 0.2% of full scale)
Repeatability + 0.15% of full scale + 0.15% of full scale
Maximum Working Pressure 500 PSIG 500 PSIG
Seals Viton® (Standard), Buna-N Viton® (Standard), Buna-N
Analog I/O (Standard) 0 - 5 VDC 0 - 5 VDC
Analog I/O (Optional) 0 - 10 VDC, 0-20 mA, 4 - 20 mA 0 - 10 VDC, 0-20 mA, 4 - 20 mA
Digital I/O (Standard) RS-232, RS-485 RS-232, RS-485
Analog Connector 15-pin subminiature D 15-pin subminiature D
Dingial Connector Dual RJ-12 Dual RJ-12
Fittings Swagelok® Swagelok®
Power Requirements + 12, + 15, + 124 VDC (11 - 36 VDC)
6.7 W
+ 12, + 15, + 124 VDC (11 - 36 VDC)
6.7 W
Wetted Materials 316 SS, 302 SS, Ni 200, Viton® (Optional Buna-N®), PTFE 316 SS, 302 SS, Ni 200, Viton® (Optional Buna-N®), Kalrez, PTFE