The Hastings Instruments Digital
Metaline represents a culmination of
over 50 years of expertise in the
designing and manufacturing of reliable
high-quality massflow instrumentation.
This digital instrument complements an
extensive array of wide-ranged analogue
instrumentation.
As with most all Hastings flow instruments,
this device is based on a
modular design. The heart of the
instrument is a patented thermoelectric
sensor and shunt assembly which
provides a highly linear response over a
broad range of flow measurement. The
physics of the design is elegant. The
inboard sensor taps sample the laminar
flow in the annular passageway formed
by the shunt and base. The change in
sensor power required to maintain a
constant sensor temperature is proportional
to the flow. Further, the sensor is
designed to be removable/replaceable in
the field to virtually eliminate downtime
due to clogging. For a recalibration in
the field (a calibration standard is
required), the customer needs to simply
set the zero and span pots or send the
zero and span commands via the digital
cable.
This digital instrument offers
alarms for sensor failure. All of the
known gas conversion factors are
factory programmed in the memory so
that changing gases is accomplished by
simply sending the appropriate command.
The Digital Metaline can operate
with standard analogue commands and
outputs or with digital data as received.
For further information please contact
us or download the PDF Data sheet download