| Standard
Ranges: |
0-50, 0-100 mTorr 0-1, 0-2, 0-5, 0-10,
0-100 Torr mbar & optional ranges are also available |
Ambient
Relative:
Humidity Effects: (0-95% RH) |
<0.005% of FS |
| Accuracy: |
0.15% of reading + 0.005% of FS |
Overpressure
Limit: |
45 PSIA |
| Repeatability:
|
±0.01% of FS |
Leak
Rate to Ambient: |
< I x 10-10 std cc/sec
at 760 Torr applied pressure |
| Resolution:
|
0.005% of FS |
Internal
Volume: |
8.33 cc including 1/2"
OD tube |
| Warm
up Time: |
2 hours |
Materials
Exposed to Pressure Media: |
Inconel, Monel
(See fitting options for other materials) |
| Heater
Temperature: |
655 - 45°C
658 - 100°C
659 - 123°C |
Power
Required: |
+ 15 VDC + 5% regulated
to 1%
655 - 250mA maximum
658 - 500mA maximum
659 - 700mA maximum |
| Ambient
Operating Temperature: |
655 - 15 - 40°C
658 - 15 - 70°C
659 - 15 - 70°C |
Output
Signal: |
0 - 10 VDC, linear with
pressure |
| Storage
Temperature: |
-45 to + 85°C |
Output
Impedance: |
<0.1ohms |
| Temperature
Effect on Zero: |
> I Torr FS = <0.002% of FS/°C
< I Torr FS = <0.005% of FS/°C |
Output
Noise:
(0 - 2 KHz) |
<0.002% of FS rms. |
| Temperature
Effect on Sensitivity: |
<0.02% of reading/°C |
Fittings:
(other optional fittings are available) |
1/2" OD Tube standard
NW 16 (316SS)
Cajon 8 VCR Female (316SS)
Cajon 8 VCO Female (316SS)
Tri-Clover (316SS) |
| Barometric
Pressure Effects: |
<0.005% of FS/760 Torr |
Weight:
|
1.2 lb (.55 Kg) |