Chell Instruments
 

Model 600 Barocel® Absolute Vacuum/Pressure Transducer


PDF Data sheet download (410kB)

  • Wide range
    4 decades
  • High accuracy
    0.15% of reading
  • High resolution
    0.01% full scale
  • Unaffected by overpressure
    repeats within 0.01 % full scale after being exposed to 35 psia
  • True total pressure
    reading independent of gas composition
  • High level DC output
    0-10 volts, linear with pressure
  • Superior zero stability
    patented design eliminates errors due to temperature and overpressure effects
  • Stable zero reference
  • Alternative for McLeod, Thermocouple and Pirani Gauges
  • System versatility
    compatible with a variety of readouts and controllers
  • Corrosion Resistant
    measuring cell constructed of Inconel/Monel for high corrosion resistance.
Description

The Barocel Model 600 transducer is an advanced design capacitance manometer which uses patented technology to improve zero stability after temperature excursions or pressure overloads. An integral vacuum reference, of all-welded construction, contains gettering material to ensure long-term stability.

In vacuum systems, the Model 600 is the ideal replacement for McLeod, Pirani, or thermocouple gauges. Since pressure is sensed by electrically measuring pressure-induced displacement of a thin metal membrane, calibration accuracy is unaffected by the chemical composition of input media.

Dirty or corrosive gases and radioactive gases such as tritium may be monitored without affecting accuracy because the electrically active portion of the sensor is contained in the sealed reference vacuum.

The transducer's analogue output is directly proportional to pressure and may be recorded by data acquisition systems or displayed using an Edwards readout/power supply.

Technical Data
Pressure ranges: 0-10, 0-100, 0-1000 torr 0-10, 0-1000, 0-1000 mbar Ranges to 10,000 torr also available Overpressure effect: Repeats within 0.01% full scale after 35 psia exposure
Accuracy: 0.15% of reading standard, 0.5% of reading also available Material exposed to pressure media: Inconel and Monel
Resolution: Better then 1:104 Internal volume: 8.3 cm3 including 1/2” dia. tube
Operating temperature: 0 to 65°C Time constant: 8 msec
Temperature effect on zero: Less than 0.005% full scale /°C Output signal: 0-10 VDC, linear with pressure
Temperature effect on sensitivity: Less than 0.02% reading /°C Power requirements: 15 VDC ±5% regulated to
± l % at 30mA
Overpressure without damage: 1.25 times full scale range or 35 psia, whichever is greater RFI suppression: Standard on all models
Weight: 1.5 lb (0.7 kg)    

Displays and Controllers for use with the model 600

16 bit Transducer and Controller Display Self contained displays for pressure and flow transducers / controllers.
5 Digit Linearising Transducer Display Self contained linearising displays for pressure and flow transducers.