Temperature
Measurement accuracy of temperature of the fluid in the air extraction
line is vitally
important, as the AIM uses this to calculate the partial pressure of saturated
steam.
This figure is then deducted from the absolute pressure to give the ratio
of steam to air
in the extraction line.
Assuming the Pitot tube was purchased from Chell it will have a platinum
resistance
thermometer fitted down the centre, and is supplied with a mating 4-pin
connector and
cable to connect to the AIM. The On-Site Maintenance Kit includes a battery
powered
hand held temperature indicator and platinum resistance thermometer, with
UKAS
accredited calibration. Connecting the indicator to the PRT in the Pitot
tube and comparing
its reading to that from the AIM will verify the AIM display, whilst replacing
the
PRT in the Pitot tube with the one in the kit will detect any problems
with the PRT. |